FLUORAT®-02-series Analyzers
|
|
FLUORAT®-02, which was a pioneering instrument developed by LUMEX
, has gained a wide use in many analytical laboratories. It is a
basic model for a variety of analytical fluorometric, photometric,
and chemoluminometric techniques. This versatility allows an analyst
to choose the optimal method for determination of each component
of interest. The Fluorat analyzers employ the modular principle.
The base unit, Fluorat-02, is a filter-based fluorometer, photometer,
and chemoluminometer. All these modes of operation are implemented
in the same optoelectronic scheme. In addition, the Fluorat-02-2M
version is a time-resolving fluorometer, which can be used as a
fluorometric detector in any HPLC system. The relevant measuring
cells are interchangeable. FLUORAT®-02 is an automated analyzer,
supplied with a package of certified methods for monitoring the
quality of drinking and waste water in compliance with sanitary
regulations. It can be also used for determination of a number of
air, soil, and food product constituents. The analytical methods
implemented in this system are chosen to reliably detect at least
half of the prescribed UTC and to provide the simplest sample preparation.
For example, it takes 2 to 3, 5 to 7, and 2 to 3 minutes, respectively,
to prepare a sample of petroleum hydrocarbons, phenols, and surfactants
for the analysis.
Description
|
|
- Automatic pressure or electrokinetic sample injection
- Built-in capillary rinsing system
- Convenient access to the vials and capillary
- Possibility of visual check of the position of the
vials and capillary rinsing
- Forced air cooling of the capillary in CAPEL103R/104,
water cooling system in CAPEL103RT/104T and in CAPEL-105
- Built-in monochromator (190 - 380 nm) in CAPEL-105
|
|
Wide spectral range
FLUORAT®-02 analyzer is equipped with a high-pressure xenon arc
lamp, which emits high-intensity light in a spectral range from
200 to 950 nm.
Versatility
The instrument is basically a fluorometer, however, it can be also
operated as a photometer, chemiluminometer, or phosphorimeter. The
cell compartment accommodates 10x10 mm cells for fluorometry, or
10x20 and 10x40 mm cells for photometry (implemented in the FLUORAT®-02-2M
model).
Use in the field
The analyzer is powered from AC mains or from a 12-V accumulator
battery, which allows its use in mobile laboratories.
Automated operation procedures
The names of the methods, result processing technique, and calibration
coefficients are displayed in the user's menu. The menu content
and the loaded calibrations are stored in a non-volatile memory.
The self-diagnostics procedure is automatically performed when the
instrument is switched on and during measurement. When carrying
out an analysis, an operator merely chooses the necessary method
from the menu and, having installed a cell, starts measuring. The
concentration of a component being measured is indicated on a built-in
LC display or is transmitted via a serial port to a standalone computer
for further processing.
The table lists the detection limits for some substances in the
analysis of water using FLUORAT®-02 analyzers in compliance with
the certified methods. It should be noted that these methods are
optimized for reliable identification of UTCs, and the actual detection
limits are lower than those given in the table.
Water. Detection limits (mg/l)
| PERFORMANCE CHARACTERISTICS
|
| Fluorescent
methods |
DL (mg/L) |
sample volume,
mL |
| petroleum hydrocarbons |
0.005 |
100 |
| phenols |
0.0005 |
250 |
| cationic surfactants |
0.01 |
5 |
| formaldehyde |
0.002 |
25 |
| nitrites |
0.005 |
5 |
| nitrates |
0.1 |
5 |
| sulfides |
0.005 |
10 |
| fluorides |
0.01 |
10 |
| copper |
0.005 |
2 |
| zinc |
0.005 |
50 |
| arsenic |
0.001 |
100 |
| cadmium |
0.0005 |
200 |
| selenium |
0.0001 |
100 |
| beryllium |
0.0005 |
1000 |
| stannum |
0.01 |
10 |
| boron |
0.05 |
5 |
| aluminum |
0.01 |
10 |
| cobalt |
0.0001 |
25 |
| iron |
0.05 |
10 |
| nickel |
0.01 |
100 |
| aluminum |
0.01 |
10 |
| Phosphorescent
methods |
|
|
| chromium (total) |
0.002 |
2 |
| Chemiluminescent
methods |
|
|
| uranium* |
0.0002 |
5 |
| Cryoluminescent
methods |
|
|
| benzo[a]pyrene*
|
0.000001 |
|
| lead* |
0.001 |
|
|
Notes: *) only for FLUORAT®-02-2M model
FLUORAT®-02-3M and FLUORAT®-02-2M analyzer versions
FLUORAT®-02-2M is used to measure the phosphorescence intensity,
operating in the time-resolving fluorometer mode. This mode is used
for the detection of uranium, with very high sensitivity and selectivity.
The analyzer can also operate with a fiber optic cable, which allows
measuring luminescence or reflected light from the surface of an
outside sample, such as paper, cathodoliminophors, etc. FLUORAT®-02-2M
can be combined with CRYO-1/2 attachment and can also be used as
a fluorometric detector for HPLC systems.
FLUORAT®-02-02/3M features:
- operation in a user-friendly menu mode to analyze common contaminants
- printout of the results of analysis
- storing results in the built-in memory
- power supply from AC mains or from a 12-V accumulator battery
- self-diagnostics: the device automatically identifies failures
in all units
| SPECIFICATIONS
|
| |
02-3M |
02-2M |
| Spectral range |
200 - 650 nm
|
200-650 nm |
| Cell types |
Q10, Q20, Q40
|
|
| Power requirements |
110/220, dc
12 V, 36 W |
|
| Size |
300x300x90 mm |
330x300x120
mm |
| Weight |
8 kg |
10 kg |
|
|
 |
|
|

|