  Field Transportable IRM-915 Mercury Stack Monitor

Types of analysis

RATA-proven OhioLumex IRM-915 Mercury Stack Monitor is developed for temporary at-stack installation and Flue Gas testing for “Total” or “Elementary” mercury.

IRM-915 Analyzer IRM-915 Probe

Designed as an instrumental replacement for Ontario Hydro Method, this is a good tool for testing certification of permanent CEMM’s, and on-site performance evaluation of Mercury Control Technologies.

IRM-915 Mercury Stack Monitor has recently passed RATA (Relative Accuracy Test Audit). Result are available per request.

Technology

Real-Time continuous mercury monitor is based on the analytical approach of Thermo catalytic conversion and Atomic Absorption for detection of mercury with Zeeman background correction. This approach requires no preconcentration on gold eliminating the associated problems. The use of multipath cell combined with “dry” converter provides highest sensitivity with no interferences from the combustion gas matrix. High converter temperature (700°C), short residence time and up to 1:100 dilution prevents Mercury atoms from recombining with any “active” species generated due to high temperature decomposition of flue gas.

Heated Probe and Heated Filter with Dilution/Conversion assembly are used for “High” or “Low” particulate loading. Results are reported on “wet” basis as regulation requires.
Mercury speciation is possible with this system.

Technical specifications
Detection limit for Flue Gas 0.5-1000 µg/dscm, "Total" mercury on "wet" basis
Calibration Mercury Calibration gas, NIST traceable Standard Referenced Materials with Hovacal generator
Set-up or take-down time less than 3 hours
Utilities 110v/60Hz, 20 A
Compressed air, 15-20 L at 80 psig
Dimensions and weight 1 Rolling Pelican Case, 90 lb
1 Probe/Filter/Converter, 150 lbs
Data processing MS Windows operating software, laptop computer
Data output 1 sample/min
Maintenance Particulate cartridge filter and scrubber replaced as required (weekly)
Warranties Hg Analyzer is detachable and can be repaired or replaced in 48 hours
Parts for Filter/Converter/Diluter supplied in 48 hours
 