Ohio Lumex Zeeman Mercury Analyzer








PRODUCTS Search
PRODUCTS | ACCESSORIES

 

CAPEL-series Capillary Electrophoresis Systems

Description


- Automatic pressure or electrokinetic sample injection
- Built-in capillary rinsing system
- Convenient access to the vials and capillary
- Possibility of visual monitoring of the position of the vials and capillary rinsing
- Forced air cooling of the capillary in CAPEL103R/104, liquid cooling system in CAPEL103RT/104T and in CAPEL-105
- Built-in monochromator (190 - 380 nm) in CAPEL-105


CAPEL-103R/103RT , their upgraded automated version CAPEL-104/104T and CAPEL-105 are high-performance capillary electrophoresis instruments, equipped with a photometric detector. The CAPEL-103R/103RT analyzers have manually controlled samplers with 4 inlet and 2 outlet positions, and CAPEL-104/104T/105 have autosamplers with 10 inlet and 10 outlet positions. The samples, rinsing and buffer solution are fed automatically according to a preset program. The autosampler in CAPEL-104/104T/105 increases the number of analyzed samples to 7, and the customized software makes the analysis fully automated. All instruments are equipped with a reversible polarity voltage supply units and with a customized report outputs (hard copy and file). The CAPEL-103RT/104T/105 have a liquid cooling system with temperature control for capillary cooling.

Design and operation features


Detectors
CAPEL-103R / -103RT, -104/ -104T with one wavelength UV-detector have a working wavelength of 254 nm. CAPEL-105 is equipped with variable wavelength UV-detector, with a wavelength range from 190 to 380 nm.

Vials
Standard 1.5 mL vials are used, with a minimal sample volume 100 µL.

Injection modes
Injection modes employ controlled pressure profile injection with variable peak pressure, with programmable peak pressure ranging from 0 to 30 mbar. The injection time can be programmed, too.

Electrophoresis power
Voltage range can be set from 1 to 25 kV
(manually reversible polarity)
Current settings may be varied from 0 to 200 mA.


System control
Membrane keyboard
Four string 20 characters and 30 string 16 characters LCDs in Capel-103R/103RT and Capel-104/104T/105, respectively.
Data acquisition and processing system are based on a standard PC and Chrom&Spec® software with MSWindows® user interface.

Chrom&Spec® software features
Real time electrophoregram visualization
Electrophoregram data processing
Computation of electrophoresis system parameters
Customized report output (hard copy and file), data exchanges with worksheets, databases and word processors
Extensive data processing algorithm customization
GLP and GALP compliance.
Data exchanges with worksheets, databases and word processors.


PERFORMANCE CHARACTERISTICS
Detection wavelength 254 nm (optional 214 and 229 nm)
Dynamic range of optical density measured From 2x10-5 to 2x10-2
Reproducibility of the retention time 0.5 %
Reproducibility of the peak magnitude 1 %
Voltage range settable From 1 to 25 kV
Capillary length/diameter 30 – 100 cm / 25, 50, 75, and 100 µm
Size and weight: (CAPEL-103R/103RT) 420x330x360 mm, 16/18 kg
Size and weight: (CAPEL-104/104T, 105) 420x360x440 mm, 22/24 kg
Power requirements 110/220 V, 150 W

For additional information on features and applications of the system, please refer to our presentation (format *.PDF): "HPCE with CAPEL systems. Àpplications."  (Download zip-file 270 Kb).





HOME | ABOUT COMPANY | PRODUCTS | APPLICATIONS | ANALYSIS TECHNIQUE | NEWS | CONTACT
 
Copyright © 2001 - 2006 OhioLumex Co. Design NDA Solaris