Field Mercury Mini CEM 2x for Gas and Solid Samples
|
|
Mercury analyzer RA-915 Mini CEM 2x is designed for field
on-site monitoring of Flue Gas for total mercury and analysis
of solid samples. This instrument analyzes samples of coal, carbon
or fly ash and is ideal for on-site performance evaluation of
Mercury Control Technologies.
| Mercury
Analyzer RA-915 Mini CEM 2x |
|
Real-Time continuous monitor for "total" mercury RA-915
Mini CEM is based on the analytical approach of Thermo catalytic conversion
and Cold Vapor Atomic Absorption for detection of mercury with
Zeeman background correction. This approach allows operation with
no preconcentration on gold eliminating the associated problems.
The use of multipath cell combined with "dry" converter provides
highest sensitivity with minimal interferences from the combustion
gas matrix. Mercury measurements take place in the heated cell
zone of converter coupled to spectrometer. This prevents Mercury
atoms from recombining with any species generated due to high
temperature decomposition of flue gas. Measurement results are
reported on the "Dry" basis, no recalculation for "moisture content"
required. Also, measurements are "flow independent" in wide range.
Only particulate filter and heated transfer lines required for
sample conditioning system. Elemental mercury could be measured
by bypassing converter introducing of flue gas directly to analyzer.
Modular and rack mounted RA-915 Mini CEM units are presented on the pictures.
| Technical Specifications
|
| Detection limit for
Flue Gas |
0.5 µg/dscm, "total"
mercury |
| Direct analysis (results
in two minutes) of sorbent/coal samples, detection limit |
0.5 µg/kg (ppb)"total"
mercury |
| Calibration |
Multi point calibration, NIST
traceable Standard Referenced Materials |
| Set-up time |
less than one hour |
| Utilities |
110v/60Hz, 1000Watt |
| Dimensions and weight |
19"x11"x23", 50
lb |
| Data processing |
MS Windows operating software |
|
|
 |