VOC Monitoring System (Ei2300-VOC)

Precise on-site measurement of Volatile Organic Compounds (VOCs) in biogas, renewable natural gas (RNG), and natural gas (NG)

Available for Download

The VOC Continuous Monitoring System is an independent and fully automated measurement system which has been developed to precisely quantify the concentration of ionizable Volatile Organic Compounds (VOCs), in biogas, Renewable Natural Gas (RNG) and Natural Gas. The instrument separates compounds in two independent stages, combining the selectivity of a Gas Chromatograph with the extraordinary sensitivity (low ppbv) of an Ion Mobility Spectrometer (IMS).  Two applications are available.

VOC Filter Breakthrough Monitoring

Rather than analyzing hundreds of single compounds by defining evaluation areas on the fixed position of respective compounds, we widened the evaluation area and performed a grid-based analysis that sums up all ionizable VOCs, then monitors the change.

After installing fresh filter media, the first measurements are set as the baseline for a new filter.  The sum of individual changes is determined for consecutive measurements.  Upon a breakthrough of the filter, the sum will start to increase with elevated compounds and respective concentrations. This will allow operators to monitor the state of the VOC filter media, all at a reasonable cost.

VOC Compound Monitoring

Need to monitor specific VOCs?  No problem.  Similar to our approach for monitoring nine individual Siloxane compounds, the Ei2300-VOC can be calibrated to measure specific VOC compounds based on your monitoring requirements.

All options are available in configurations suitable for either general purpose or hazardous areas, and can be located indoors or outdoors.

Product Details

VOC Monitoring System (Ei2300-VOC) Features

  • Two-fold Matrix Separation
  • Gas Chromatograph
  • Ion Mobility Spectrometer (IMS)
  • Precise Measurements
  • Easy to Operate
  • No Analytical Specialist Needed

Options

Hazardous Rated Enclosure
  • Purged
  • Class 1, Division 2
Side Mounted Air Conditioning Unit
  • Standard
  • Class 1, Division 2
Remote Access Hardware
  • Cellular Service
  • Secure Cradle Point

Configurations

19″ Rack, Wall Mount

VOC Monitoring System (Ei2300-VOC) Technical Specifications

Specifications
Separation I GC Retention Time
Gas Chromatopgraph (GC) Isotherm (< 80° C)
Capillary Column 30m
Separation II Drift Time
Ion Mobility Spectrometer (TOF-IMS)
Ionization β-radiation – 3H
Activity 300 MBq, below the exemption limit of 1 GBq acc. To EURATOM guideline
Detection Electrometer,
Ion Mobility Spectrometer
LoD for VOCs One digit ppb
Pneumatics Two Electronic Pressure Controllers (EPC) for flow stability and GC ramping
Sampling Ambient pressure, Firmware controlled pump plus heated electrical six port valve, Liquid trap (water condensation) optional
Purge Gas Requirement Nitrogen: Quality 5.0 (cylinder or generator)
Operation Manual: 6.4" TFT touchscreen
Online Remote Test: User defined intervals
Data Storage (Internal) 16 GB Flash Memory
Data Output USB, 4-20mA, MODBUS (TCP)
Ambient Temperature Operation 0 - 40° C
Moisture Up to 95% non-condensing
Power Range 100 - 240 V AC, 50 - 60 Hz
Power Consumption < 200 Watt
Weight 34 lbs (15.5 kg)
Dimensions 24 x 30 x 12 in. (Wall Mount)
17.7 x 17.1 x 7 in. (Rack Mount)
Housing 19:, IP 20 Enclosure, CE Marking

Typical Applications for VOC Monitoring System (Ei2300-VOC)

  • Wastewater Treatment Plants
  • RNG Interconnection Rules Compliance
  • Landfills
  • Dairy Farms
  • Hog and Pig Farms
  • Food Waste Digesters (co-digestion)
  • Others

Available Measurements for VOC Monitoring System (Ei2300-VOC)

  • Volatile Organic Compounds
  • Options:
    1. Calibrate for specific VOCs
    2. Monitor change in total VOCs for filter media breakthrough